Tender are invited for 1 unit Cardiotocography Machine for BAIS (N-2272-2024)(H) 2024-31
Procurement Mode : Negotiated Procurement - Small Value Procurement (Sec. 53.9)
Classification : Goods
Category : Hospital / Medical Equipment
Contract Duration : 0 Day/s
Budget for the Contract : PHP 65, 000.00
Area of Delivery : Negros Oriental
Description : 1 Cardiotocography Machine 1 unit 65, 000.00
Specificatons:
1. Capable of external monitoring of fetal heart rate (FHR) and maternal uterine activity (UA).
2. Alphanumeric display shows FHR1, FHR2 and UCs and alarms.
3. Automatically detects transducers when they are plugged in.
4. Includes remote switch for event marking.
5. Automatic self-test on power up.
6. System reports with status and alarms.
7. Unit includes a battery allowing for 2 hours continuous operation.
8. Power requirements: 100 - 240 Volts - 50/60 Hz.
FETAL HEART RATE MONITORING
1. Fetal heart rate detected by ultrasonic transducer/probe.
2. Capable of measuring FHR in the 50-240 beats per minute (bpm) range with 1 bpm resolution and 2 bpm accuracy.
3. Can monitor twins (i.e., has two transducers).
MATERNAL UTERINE ACTIVITY MONITORING
1. Capable of measuring relative uterine contractions (UCs) in the range of 0-100 units with at least 1 unit resolution and 1 unit accuracy.
PRINTER/RECORDER
1. Integrated thermal printer with automatic and manual print-out ...