SOUTHERN TAGALOG REGIONAL HOSPITAL has floated a tender for Calibration and Preventive Maintenance of Infant Radiant Warmer. The project location is Philippines and the tender is closing on 16 Apr 2024. The tender notice number is SV-2024-025, while the TOT Ref Number is 99795111. Bidders can have further information about the Tender and can request the complete Tender document by Registering on the site.

Expired Tender

Procurement Summary

Country : Philippines

Summary : Calibration and Preventive Maintenance of Infant Radiant Warmer

Deadline : 16 Apr 2024

Other Information

Notice Type : Tender

TOT Ref.No.: 99795111

Document Ref. No. : SV-2024-025

Competition : ICB

Financier : Self Financed

Purchaser Ownership : Public

Tender Value : PHP 5500.00

Purchaser's Detail

Name :Login to see tender_details

Address : Login to see tender_details

Email : Login to see tender_details

Login to see details

Tender Details

Tender are invited for Calibration and Preventive Maintenance of Infant Radiant Warmer

Procurement Mode : Negotiated Procurement - Small Value Procurement (Sec. 53.9)

Classification : Goods

Category : Hospital / Medical Equipment Services

Contract Duration : 0 Day/s

Budget for the Contract : PHP 5, 500.00

Area of Delivery : Cavite

Description : 1 Unit for Calibration and Preventive Maintenance of Infant Radiant Warmer



Infant Radiant Warmer

a. Frequency: SEMI-ANNUAL

b. Quantitative Test (Calibration Parameter):

(Calibration must include but not limited to the ff test)

b.1 Temperature Mapping

b.2 Electrical Safety Test



Brand/Model: Novos/KR1000

1. KR-11188-002



Standard Requirements:

a. Supplier must provide Service and Test Report for each unit as compliance for the requirement of STRH. Details of Calibration Tools such as brand, model, serial number, and calibration date range must be indicated in either in Service Report or Test Report.

b. Supplier must provide a copy of Calibration Report or Certificate for the Calibration Tools used (For Temperature Data Logger and Electrical safety Analyzer)

c. Supplier must provide a calibration sticker per unit indicating the start and end date of calibration.

d. Supplier must conduct testing and calibration of unit for at least three (3) points per parameter.

e. Supplier must recommend all necessary services and parts needed for functionali...