SOUTHERN TAGALOG REGIONAL HOSPITAL has floated a tender for Calibration and Preventive Maintenance of Infant Radiant Warmer. The project location is Philippines and the tender is closing on 29 Apr 2024. The tender notice number is SV-2024-025A, while the TOT Ref Number is 100298607. Bidders can have further information about the Tender and can request the complete Tender document by Registering on the site.

Expired Tender

Procurement Summary

Country : Philippines

Summary : Calibration and Preventive Maintenance of Infant Radiant Warmer

Deadline : 29 Apr 2024

Other Information

Notice Type : Tender

TOT Ref.No.: 100298607

Document Ref. No. : SV-2024-025A

Competition : ICB

Financier : Self Financed

Purchaser Ownership : Public

Tender Value : PHP 5500.00

Purchaser's Detail

Name :Login to see tender_details

Address : Login to see tender_details

Email : Login to see tender_details

Login to see details

Tender Details

Tender are invited for Calibration and Preventive Maintenance of Infant Radiant Warmer

Procurement Mode : Negotiated Procurement - Small Value Procurement (Sec. 53.9)

Classification : Goods

Category : Hospital / Medical Equipment Services

Contract Duration : 0 Day/s

Budget for the Contract : PHP 5, 500.00

Area of Delivery : Cavite

Description : Calibration and Preventive Maintenance of Infant

Radiant Warmer

Infant Radiant Warmer

a. Frequency: SEMI-ANNUAL

b. Quantitative Test (Calibration Parameter):

(Calibration must include but not limited to

the ff test)

b.1 Temperature Mapping

b.2 Electrical Safety Test

Brand/Model: Novos/KR1000

1. KR-11188-002

Standard Requirements:

a. Supplier must provide Service and Test

Report for each unit as compliance for the

requirement of STRH. Details of Calibration

Tools such as brand, model, serial number,

and calibration date range must be indicated

in either in Service Report or Test Report.

b. Supplier must provide a copy of Calibration

Report or Certificate for the Calibration

Tools used (For Temperature Data Logger

and Electrical safety Analyzer)

c. Supplier must provide a calibration sticker

per unit indicating the start and end date of

calibration.

d. Supplier must conduct testing and calibration

of unit for at least three (3) points per

parameter.

e. Supplier must recommend all necessa...

Documents

 Tender Notice