Fourier Transform Ion Cyclotron Resonance Highresolution Mass Spectrometry System... Tender

NATIONAL UNIVERSITY CORPORATION has floated a tender for Fourier Transform Ion Cyclotron Resonance Highresolution Mass Spectrometry System 1 Set. The project location is Japan and the tender is closing on 11 Apr 2022. The tender notice number is , while the TOT Ref Number is 64222171. Bidders can have further information about the Tender and can request the complete Tender document by Registering on the site.

Expired Tender

Procurement Summary

Country : Japan

Summary : Fourier Transform Ion Cyclotron Resonance Highresolution Mass Spectrometry System 1 Set

Deadline : 11 Apr 2022

Other Information

Notice Type : Tender

TOT Ref.No.: 64222171

Document Ref. No. :

Competition : ICB

Financier : Self Financed

Purchaser Ownership : Public

Tender Value : Refer Document

Purchaser's Detail

Name :Login to see tender_details

Address : Login to see tender_details

Email : Login to see tender_details

Login to see details

Tender Details

Fourier transform ion cyclotron resonance high-resolution mass spectrometry system 1 Set

Classification :

0024 Professional/Scientific & Controlling Instruments & Apparatus

Summay of Notice :

⑴ Classification of the products to be procured : 24

⑵ Nature and quantity of the products to be purchased : Fourier transform ion cyclotron resonance high-resolution mass spectrometry system 1 Set

⑶ Type of the procurement : purchase

⑷ Basic requirements of the procurement :

A Fourier transform ion cyclotron resonance mass spectrometry

a Superconducting magnet should be equipped with the following functionalities

・Magnetic field should be higher than 12 T.

・The refrigerant of the superconducting magnet shall be liquid helium only.

b Cryo cooling system

・To suppress the evaporation of liquid helium inside the magnet, it must have a cooling system.

c Chiller system for cryo cooling

・It shall have a water-cooled chiller system for cryo-cooling.

d Console system

・Has the function of converting frequency information detected from the ICR cell into m/z spectral information.

e Mass spectrometry system

・The ICR cell detector must be cylindrical.

・Ion selection quadrupole, quadrupole collision cell, and hexode ion transfer mechanism in front of the ICR cell.

・The ion transport section shall have an ion funnel mechanism.

・The measured mass range must be m/z 100-10, ...

Documents

 Tender Notice


Procurement Documents for Japan

Access a comprehensive library of standard procurement documents specific to Japan. Here, you'll find all the essential forms, guidelines, and templates required for tender applications and submissions in Japan

Explore Procurement Documents for Japan


Want To Bid in This Tender?

Get Local Agent Support in Japan and 60 More Countries.

View All The Services


View Tenders By


Publish Tenders


Have Any Dispute With The Purchaser?