Procurement Summary
Country : Iraq
Summary : Supply of Monitor for Sever Patient, Infusion Pumps, Large-Volume, Infusion Pumps, Syringe, Cardiac Ultrasonic Scanning Systems (Echo), Portable, etc
Deadline : 14 Feb 2023
Other Information
Notice Type : Tender
TOT Ref.No.: 78097924
Document Ref. No. : 55/2023/16
Financier : Self Financed
Purchaser Ownership : Public
Tender Value : Refer Document
Purchaser's Detail
Name :Login to see tender_details
Address : Login to see tender_details
Email : Login to see tender_details
Login to see detailsTender Details
Tenders are invited for Supply of Monitor for Sever Patient, Infusion Pumps, Large-Volume, Infusion Pumps, Syringe, Cardiac Ultrasonic Scanning Systems (Echo), Portable, etc
Monitor for Sever Patient(with Picco), Infusion Pump, Large-Volume, Infusion Pump, Syringe, Cardiac Ultrasonic Scanning System (Echo) Portable, Defibrillator, Ecg, Radiographic Units, Mobile (Dr), Blood and Fluid Warmer, Etc.
Bidding documents for medical & service equipment purchasing fees: ($ 500) (five hundred
USD).
1-Device Name: Bed Side Monitor For Sever Patient (With Picco)
Application : Basic & Advanced Physiologic Parameters .
Acuity Area :Icu, Rcu, Picu, Cardiac Post-Operative Care Unit, Renal Transplant
Post-Operative Care Unit.
Patient Type Adult, Pediatric (With All Accessories Needed)
Nibp Yes Systolic, Mean, And Diastolic Pressures
Ecg Yes ≥ 5 Leads
Parameters Analyzed St Segment & Arrhythmia Monitoring Facility.
Ibp
Labels
Yes ≥2 Channels
Cvp, Others To Be Specified.
Hemodynamics, & Cardiac Output(Picco)
Svo2
Yes With All Accessories Needed
Optional
Spo2 (Reusable Probe)( Measuring Range: Spo2: 1 To
100%)
Yes
Temperature (°C) Yes
Capnography, Etco2 Yes
Pulse Rate Yes
Respiratory Yes
Arrhythmia Analysis+ St Analysis
No. Of Leads Analyzed
Yes
≥6
Trending Graphical
Tabular
Length Of Time, Hr ≥ 24 Hr.
Alarms
For All Monitored Parameters
Audible And Visual.
Alarms: User-Selectable Upper And Lower Limits
Arrhythmia Alarm81
Display
Size
Traces
Lcd Or Led, Or Tft Color Touchscreen
≥ 15 In.
≥ 6
Connectable To Central Station Monitor Yes
Graphical And Numeric Monitored Data Should Be
Recordable
Yes ≥ 24 Hr.
Networking
Interface
Auxiliary Output
Connection
Yes
Yes
Digital
Wired Or Wireless
Recorder Optional
Equipment Alarms Cuff Leak, Cuff Disconnect, Hose Leak, Low
Battery, Low Systolic Bp, High Systolic Bp,
Low Diastolic Bp, High Diastolic Bp, Low Mean
Bp, High Heart Rate, Low Heart Rate, Low Pulse
Oximetry, Low Respiratory Rate, And High
Respiratory Rate, Other.
Orientation Wall Mount, Or Bed Side Or With Trolley
Battery(Operating Time) Yes ≥1hr
Lead Off Status Yes
Power Requirement Power Input To Be 220-240vac, 50hz, 3 Pin
Plug (G Type).
Environmental Requirements: The Supplier Should Be Confirmed That: The
Equipment Is Suitable For Work In Climate
Conditions Of Iraq In Terms Of Temperature And
Humidity.82
Self-Test Mode Yes
Activation Indicators
Visual Yes
Audible Yes
Volume Control Yes
Power Setting Displayed Yes
Cooling Convection Or Fan
Standard Accessories (Reusable) Patient Electrode .
Connecting Cable For Patient Electrode.
Electrode Holder With Cable.
Connecting Cable For Bipolar Forceps
Electrodes (Reusable) Electrode Set (10 Different Kinds/Set).
Bipolar Forceps, Straight.
Bipolar Forceps, Bent.
Power & Environmental Requirements Power Input To Be 220-240vac, 50hz.
The Supplier Should Be Confirmed That:
The Equipment Is Suitable For Work In
Climate Conditions Of Iraq In Terms Of
Temperature And Humidity.
12-Device Name: Ultrasonic Nebulizer
Method Of Nebulization Ultrasonic Vibration
Ultrasonic Frequency ≥120khz±10%
Maximum Pressure ≤200kpa
Continuous Working Time ≥ 4 Hours
Particle Size Diameter (Microns) ≥0.5 (Mmad:3.8)
Nebulizing Rate ≥ 0.25ml/Min
Medication Capacity ≤ 20ml
Accessories Portable Battery Pack, Pediatric Mask, Adult
Mask, Ac Adapter, Air Filter And Mouthpiece,
Carrying Bag
Timer And Air Flow Adjuster Yes
Power Supply
Battery
200/240 Vac 50/60 Hz, 9-12vdc
Yes (Rechargeable ≥ 3 Hours Backup Battery)83
Charger
Typical Solid State, Over - Current Circuit
Protection
Other Specification Silent Operation (Totally Noiseless), Compact
Design And One Hand Use.
Safe - Low Battery Indicator.
Overheat Protection.
Environmental Condition The Equipment Is Suitable For Working In The
Climate Of Iraq In Terms Of Temperature And
Humidity.
13-Device Name: Examination Light, Mobile Floor Stand
Application To Provide Focused Light Directly On The Area
Where The Examination Is Being Performed.
Where Used Wards, Physician Offices & May Be Used
During Minor Surgical Procedures.
Light Head Characteristics ≤ 14"
Light Intensity (Lux) ≥ 40, 000 Lux At 1m
Fixed Focus Yes
Depth Of Field ≥ 37 Inches
Color Temperature ≤ 5, 000° K
Bulb Type Led
Bulb Life ≥ 30, 000 Hr
Bulb Replacement Manual
Camera Capability No
Articulating Arm Yes
Sterilizable Handle Yes
Color Rendering Index 90
Configuration Mobile Floor Stand With Locking Casters
Floor Stand Height ≥ 1 M.
Power Supply 220 ~240 Vac, 50/60 Hz (Three Pin Plug)
Environmental Requirements The Equipment Suitable For Working In The
Climate Conditions Of Iraq In Terms Of
Temperature & Humidity
14-Device Name: Electric Bed
Type Electric Or Electrohydraulic
Patient Controls
Location
Functions
Pendant Or Side Rail Controller
Backrest, Knee, High/Low, Others84
Nurse Controls
Patient Control Lockouts Yes
Cpr Control Yes
Maximum Patient
Weight, Kg
≥180
Siderail Length
(Percentage Of Overall)
≥1/2
Casters
Diameter, Cm
Functionsh
12.5 Cm
Brake, Steer
Removable Headboard Yes
Iv Pole
Mounts
Yes
Overall Dimensions L x W, Cm : ≥210x 90
Mattress - Bacteria And Fluid Resistant, Flameretardant, Anti-Fungal, Anti-Static,
And Non-Allergenic
- Latex Free
- Anti-Ulcer
Low Position Floor To Top Of Deck 40.01 Cm
Sleep Deck
Width
Normal Length
89.53 Cm
(203.20 Cm85
Max Head Of Bed Elevation 65°
Max Knee Elevation 20°
Max Trend And Reverse Trend 16°
Automatic Battery Backup Yes
Frame Mild Steel Or Coated Metal
15-Device Name: Hemodialysis Units
Model Basic Hemodialysis Units
Dialysate Delivery
Comfort Control, °C
Temperature Alarm Limits, °C
Conductivity Range, Ms/Cm
Flow, Ml/Min
Transmembrane Pressure, Mm Hg
≥ 30 To ≤ 40
≥ 41
≥ 12 To ≤ 17
≥ 100 To ≤ 1200
≥ -300 To ≤ 700, Alarm If 0
rx dialysis
bicarbonate
sodium therapy
ultrafiltration removal rate, l/hr
bypass indicator
yes
yes
accuracy at ±10%
yes
blood circuit
arterial pressure, mm hg
venous pressure, mm hg
blood pump range, ml/min
heparin pump range, ml/hr
≥ -700 to ≤ 800
≥ 0 to 800
≥ 0 to ≥ 500
≥ 0 to ≤ 10
disinfection
method heat and chemical or better
display type ≥ 12" inch touch screen
displayed parameters
transmembrane pressure
conductivity
flow rate
elapsed time
remaining time
prescribed time
other
yes
yes
yes
yes
yes
yes
bp reading, kt/v(measured preferable), venous and arterial
pressure.
system
interface rs232 or ethernet or usb
alarm actions
loss-of-water alarm
blood-leak detector
air/foam detector
conductivity
arterial/venous pressure
temperature, high/low
yes
yes
yes
yes
yes
yes86
other specifications - power-failure alarm and battery backup.
- accurate dialysis fluid conductivity and tmp
monitor.
- hdf
- adjustable dialysis fluid flow.
- double and single needle dialysis applications.
- card reader for patient data.
- decrease or stop uf in case of hypotension automatically.
power supply - 220vac, 50/60 hz
- three pin plug (g type)
- back up battery or ups for ≥ 15 minute
environmental requirements the supplier confirm that: the equipment suitable for work in
the climate conditions in iraq in terms of temperature &
humidity.
16-device name: operation table, general surgery
application provide an elevated surface that supports the
patient's body during general surgery procedures
table positions(degrees from horizontal)
trendelenburg
reverse trendelenburg
lateral tilt
park bench position
lateral positon
≥25
≥25
≥18
yes
yes
vertical range, cm ≥ (52- 90)
table sections ≥4
degrees from horizontal
back section
up
down
from 0 to ≥ 25
from 0 to ≥ -1087
foot/leg section
up
down
head section
from 0 to ≥ +15
from 0 to ≥ -40
variable
accessory siderail yes
perineal cutout yes
maximum patient capacity, kg ≥ 250 static
controls
remote control
type
manual override or electrical
foot/hand
yes
electromechanical, other to be specified
yes
radiolucent tabletop
c-arm accessible
yes
yes
accessories
armboard, pair of legs shaped support, sacral support
gel pad, heel & head protector, arm & heel support, leg
shaped support, other accessories should be specified
dimensions, w x l, cm
fully extended
≥ (50 x 195) or more
column housing stainless steel88
caster lock yes
power requirements
battery
power input 220-240vac, 50hz.
yes
environmental requirement the supplier should be confirmed that: the equipment
is suitable for work in climate conditions of iraq in
terms of temperature and humidity.
17-device name: operating room patient monitor
acuity area operating rooms
application measure and display waveforms and numerical data for various
parameters such as ( ecg, nibp, …..etc.) for patients.
patient type adult, pediatric, neonate(with all accessories)
beside or transport yes
modular/ configured modular or configured
parameters monitored ecg, etco2, ibp, nibp, respiration, spo2, temperature
b i s or entropy yes
etco2 with capnography yes
pulse oximetery adult & pediatric probe (reusable)
ecg
number of leads
12-lead interpretation
arrhythmia detect
no. of leads analyzed
st analysis
no. of leads analyzed
auto lead switch
yes
≤12
yes
yes
≤12
yes
≤12
yes
respiration
method impedance or etco2
ibp
no. of channels
hemodynamics, & cardiac output
yes with (reusable probe)
≤2
with all accessories needed
nibp
cuff size
temperature
adult, pediatric, neonate
yes89
no. of inputs
probe type
full disclosure
length of time (hr)
trending
length of time( hr)
≤ 4
skin or others.
yes
24
yes
≥24
alarms default setting low systolic bp, high systolic bp, low diastolic bp, high
diastolic bp, low mean bp, high heart rate, low heart rate, low
pulse oximetry, low respiratory rate, and high respiratory rate.
networking yes
display lcd or led or tft
size (inch) ≥ 10
traces ≥6
recorder optiona
Documents
Tender Notice